Map Recent Data Charts Control Radar Alarms Notes Maintenance Contact

To carry out maintenance, use the Android app.

This page is for reading and downloading what has already been recorded. The visit itself is done on a phone at the site, offline, and uploaded on the trip or afterwards (depending on cell service).

Download the app

‹ all sites

Lynch Cir 2

2026-09-05 18:56

Group
CoG
Site
Lynch Cir 2
Data file
Lynch_Cir_2.csv
Started
2026-09-05 18:56:11
Ended
2026-09-05 19:02:01
Times
phone local time UTC-04:00
Rows
82

Photographs (8)

Lynch_Cir_2_20260905_1856_dtec_185653.jpg
Lynch_Cir_2_20260905_1856_dtec_185653.jpg
Lynch_Cir_2_20260905_1856_final-dtec_190138.jpg
Lynch_Cir_2_20260905_1856_final-dtec_190138.jpg
Lynch_Cir_2_20260905_1856_final-radar_190142.jpg
Lynch_Cir_2_20260905_1856_final-radar_190142.jpg
Lynch_Cir_2_20260905_1856_final-turb_190146.jpg
Lynch_Cir_2_20260905_1856_final-turb_190146.jpg
Lynch_Cir_2_20260905_1856_final-wide_190152.jpg
Lynch_Cir_2_20260905_1856_final-wide_190152.jpg
Lynch_Cir_2_20260905_1856_radar_185928.jpg
Lynch_Cir_2_20260905_1856_radar_185928.jpg
Lynch_Cir_2_20260905_1856_turb_185944.jpg
Lynch_Cir_2_20260905_1856_turb_185944.jpg
Lynch_Cir_2_20260905_1856_wide_185617.jpg
Lynch_Cir_2_20260905_1856_wide_185617.jpg

The record

TimeEventItemValueUnit PointStandardNote
18:56:11 visit started Lynch Cir 2 CoG
18:56:11 method chosen DTEC — depth mrcm
18:56:11 job added by hand DTEC — depth not in this site's data
18:56:11 method chosen DTEC — EC ccc
18:56:11 job added by hand DTEC — EC not in this site's data
18:56:11 method chosen DTEC — temperature mrcm
18:56:11 job added by hand DTEC — temperature not in this site's data
18:56:11 method chosen Radar — depth distance
18:56:11 method chosen Turbidity — NTU ccc
18:56:11 job added by hand Turbidity — NTU not in this site's data
18:56:11 method chosen Battery maintenance swap
18:56:11 method chosen Manual data download card
18:56:12 site check Radar — no rainfall ok
18:56:12 site check Access — safe and stable ok
18:56:13 site check No damage in work area ok
18:56:17 photograph, wide photos/Lynch_Cir_2_20260905_1856_wide_185617.jpg 43.5366 -80.2553 gps
18:56:53 photograph, sensor photos/Lynch_Cir_2_20260905_1856_dtec_185653.jpg no gps
18:56:55 depth-pressure step Manual depth invert to water surface 200 mm 1 200 before the clean
18:56:58 depth-pressure step Sensor-reported depth at that moment 200 mm 1 200 before the clean
18:57:00 depth-pressure step Probe removed from water done 1
18:57:02 depth-pressure step Probe body cleaned done 1
18:57:03 depth-pressure step Probe returned to water at the same height done 1
18:57:25 taken out of order DTEC — depth left unfinished skipped manual depth after replacing
18:57:25 depth-pressure step Sensor-reported depth after replacing 0 mm 1 after the clean
18:57:40 ec precln Reading 1 50 µS/cm 1 50
18:57:52 ec precln Reading 2 50 µS/cm 1 50
18:57:54 di rinse transfer rinsed and moved to point 2 done 1 before the clean
18:58:29 ec precln Reading 1 50 µS/cm 2 50
18:58:45 ec cleaned sensor cleaned done the second pass begins here
18:59:03 temp step Reference thermometer 20 °C 1 20 before the clean
18:59:12 temp step Sensor temperature 20 °C 1 20 before the clean
18:59:14 temp step Probe lifted out of the water done 1
18:59:14 temp step Probe body and bead wiped done 1
18:59:15 temp step Probe returned to the same depth done 1
18:59:16 temp step Reference thermometer (after) 20 °C 1 20 after the clean
18:59:18 temp step Sensor temperature (after) 20 °C 1 20 after the clean
18:59:28 photograph, sensor photos/Lynch_Cir_2_20260905_1856_radar_185928.jpg no gps
18:59:31 depth-radar step Distance mount face to water surface 2025 mm 1 2025
18:59:34 depth-radar step Radar angle from horizontal 75 ° 1
18:59:37 depth-radar step Sensor-reported distance at that moment 5000 mm 1 2025
18:59:40 depth-radar step Mount and bracket secure done 1
18:59:44 photograph, sensor photos/Lynch_Cir_2_20260905_1856_turb_185944.jpg no gps
19:00:13 turbidity precln Reading 1 · LED ON 20 NTU 1 20
19:00:15 turbidity precln Reading 1 · LED OFF 0 NTU 1 20
19:00:30 turbidity precln Reading 2 · LED ON 20 NTU 1 20
19:00:32 turbidity precln Reading 2 · LED OFF 0 NTU 1 20
19:00:33 di rinse transfer rinsed and moved to point 2 done 1 before the clean
19:00:38 turbidity precln Reading 1 · LED ON 200 NTU 2 200
19:00:39 turbidity precln Reading 1 · LED OFF 0 NTU 2 200
19:00:41 taken out of order Reading 2 · LED ON at point 2 before the clean skipped the ten second wait between readings
19:00:44 turbidity precln Reading 2 · LED ON 200 NTU 2 200
19:00:46 turbidity precln Reading 2 · LED OFF 0 NTU 2 200
19:00:49 turbidity cleaned sensor cleaned done the second pass begins here
19:00:51 turbidity postcln Reading 1 · LED ON 20 NTU 1 20
19:00:53 turbidity postcln Reading 1 · LED OFF 0 NTU 1 20
19:00:55 taken out of order Reading 2 · LED ON at point 1 after the clean skipped the ten second wait between readings
19:00:56 turbidity postcln Reading 2 · LED ON 200 NTU 1 20
19:00:56 turbidity drift LED ON after minus before 90 NTU 1 20 over the ±5 limit
19:01:00 turbidity postcln Reading 2 · LED OFF 0 NTU 1 20
19:01:00 turbidity drift LED OFF after minus before 0 NTU 1 20
19:01:01 di rinse transfer rinsed and moved to point 2 done 1 after the clean
19:01:06 turbidity postcln Reading 1 · LED ON 2000 NTU 2 200
19:01:08 turbidity postcln Reading 1 · LED OFF 0 NTU 2 200
19:01:09 taken out of order Reading 2 · LED ON at point 2 after the clean skipped the ten second wait between readings
19:01:10 turbidity postcln Reading 2 · LED ON 2000 NTU 2 200
19:01:10 turbidity drift LED ON after minus before 1800 NTU 2 200 over the ±5 limit
19:01:12 turbidity postcln Reading 2 · LED OFF 0 NTU 2 200
19:01:12 turbidity drift LED OFF after minus before 0 NTU 2 200
19:01:14 turbidity remounted sensor rinsed and returned to its mount done
19:01:18 battery step Batteries changed done
19:01:18 batteries changed battery pack changed
19:01:20 download step Card removed from logger done
19:01:21 download step Files copied and verified done
19:01:21 download step Card returned and logger restarted done
19:01:38 photograph, final sensor photos/Lynch_Cir_2_20260905_1856_final-dtec_190138.jpg no gps
19:01:42 photograph, final sensor photos/Lynch_Cir_2_20260905_1856_final-radar_190142.jpg 43.5366 -80.2553 gps
19:01:46 photograph, final sensor photos/Lynch_Cir_2_20260905_1856_final-turb_190146.jpg no gps
19:01:52 photograph, final wide photos/Lynch_Cir_2_20260905_1856_final-wide_190152.jpg 43.5367 -80.2553 gps
19:01:57 final check battery 3350 mV fail
19:01:57 final check last logged 3 min ago pass
19:02:00 taken out of order ending the visit skipped DTEC — EC is not finished; the final check did not pass
19:02:01 visit ended Lynch Cir 2